16.7.1 Required Tools and Equipment

Service Module (not pictured)

Laptop computer (not pictured)

New NEG Pump

1/4” Allen wrench

Torque wrench set for 70 inch/pounds

Gloves (cotton or latex)

Lint free wipes

Methanol

MicroDuster® III (containing 1,1,1,2 tetrafluoroethane)

CAUTION

Refer to the MSDS for Personal Protection Equipment
(PPE) guidelines, wipe all tools with methanol before
starting this procedure.

Figure 16-18 Required Tools and Equipment