HAPSITE Smart Plus Operating Manual
5.4.1 Quick Reference SOP — Running Analyze (GC/MS) Loop Method
1
Verify that the Sample Loop is installed.
2
Power on or take the HAPSITE out of Extended Standby.
3
The HAPSITE will begin to prepare the start up Loop method. See
Selecting a New Method, on page 5-7
if the start up method is not the desired
method.
4
Double-click the Plus IQ icon to open the software.
5
If the HAPSITE runs an AutoTune, press OK when the screen prompts.
6
Click the RUN METHOD icon and choose the method that has been prepared.
Refer to Steps 2-4 of
.
7
Click the Run button in the pop-up window of the software
8
When the screen prompts, Collect Sample Now, hold the probe over the
sample for the specified amount of time.
9
Remove the Probe from the sample when the screen prompts, Sampling Is
Done.
10
Let the method run to completion.
11
Review data when run is complete.