7.6.1 Adjusting Base Peak Gain
1
Double-click on the Plus IQ icon. See
.
Figure 7-27 Plus IQ Icon
2
Double-click on the Tune icon. See
NOTE: Manual Tune can only be opened in Advanced User Mode.
Figure 7-28 Manual Tune Icon
3
Select default.tun file and press OK. See
Figure 7-29 Select default.tun File
4
Wait for the automated process of opening the tune to complete.
5
Check the Base Peak Gain (BPG). Is the value between 0.8 and 2.0? See
and
Figure 7-31
.