HAPSITE Smart Plus Operating Manual

Figure 12-15 Finished Analytes Chart

10d

To finish the template, the concentration of each analyte must be individually
entered if not using the global concentration flag. In this example, the
concentration of all analytes is the same, so Global is being used. The
concentration field is not a required entry. For analyte specific concentrations,
the Concentration field should contain the individual analyte concentrations for
the selected data file. Each additional file would have a specific multiplication
factor entered in the Conc/Factor field in the Data Files control panel.

Figure 12-16 Extracted Mass Peaks

10e

The Extracted Mass Peaks can also be edited to delete mass fragments with
intensities below 15%. The exception would be a molecular ion if the
fragmentation pattern is not very distinct. To delete unwanted mass fragments,
highlight the field and press the delete key. Refer to

Figure 12-16

.

11

The finished library needs to be saved. Select the Save Library button under
the Method controls. Enter a library name and save it as part of the method.
See

Figure 12-17

.