5 full scan window – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 261
7 - 19
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
7.5.5 Full Scan Window
The Full scan window (see
Figure 7-19
) is used for display purposes to evaluate
the performance of the MS. All tune adjustments should be made using the Peak
Scan windows.
Figure 7-19 Full Scan Window and Controls
Zoom . . . . . . . . . . . . . . . . . . Enables the cursor to select a section of the
peak scan window and zoom to that section.
Zoom Out. . . . . . . . . . . . . . . Returns the window to the original X axis and
Y axis scale.
Zoom Out Y Axis. . . . . . . . . Returns the Y axis to the original scale.
Placing the mouse cursor on the x axis of the full scan window and pressing the
Right Mouse Button will display the Set Scan Range Window, see
Figure 7-20
. This
allows a custom scan range to be entered and viewed.