14.6 Heated Zones

There are several heated zones that must be maintained when operating the
SituProbe to keep VOCs and moisture from condensing in the sample pathway.
The temperature zones are set through the HAPSITE Plus software and are
implemented with the start of the method. See

Figure 14-2

. The method will not

advance until the zones are heated to the correct temperatures. Once a method is
started and the zones are heating, the Transfer Line can be disconnected at the
HAPSITE without affecting the SituProbe’s heat-up process. Therefore, anytime
the SituProbe is disconnected from the HAPSITE it will maintain the heated zones
and be ready for immediate use when re-connected.

The table below contains information on each of the heated zones — their typical
operating and input range of temperatures.

Figure 14-2 Method Editor Page for Entering Temperature Settings.

Table 14-1 Heated Zone Temperatures

Heated Zone

Default Operating

Temperature (°C)

Temperature

Range

Oven

50

30 - 60

Transfer Line

60

30 - 60

Probe

40

30 - 60