HAPSITE Smart Plus Operating Manual
Figure 15-34 Yellow Protective Cover Placed Over Aluminum Storage Plug on Service Module
To complete the Service Module storage procedure, turn off the Service Module
power switch at the back left of the instrument.
CAUTION
The Service Module should be stored in a clean and dry
area. Avoid storing the Service Module in areas which are
outside the temperature range of 40 to 95 °F (5 to 35
°C).
15.11 When Power to the Service Module is Lost
In case of a sudden loss of supplied power, the Service Module will close the
HAPSITE Interconnect Plug and shut down the pumps properly. This happens
automatically in order to preserve the vacuum in the HAPSITE Mass Spectrometer
manifold and to protect the Service Module pumps from being damaged.
When power is restored, the HAPSITE and Service Module must be attached again
before the Service Module will provide proper vacuum for the HAPSITE. Refer to
section 15.7, Attaching the HAPSITE to the Service Module, on page 15-12
NOTE: See
for additional information on using the
Service Module to perform maintenance on the HAPSITE.