2.12.2.4 Understanding the Battery Charger Indicators

Each battery receptacle is associated with an indicator light which can be
illuminated in any of three colors.

Green. . . . . . . . . . . . . . . . . . . . . . . . The battery is being charged. This will

continue for 15 hours or less. If a battery with
a severely depleted charge is inserted, the
green light will flash. If it flashes for more than
10 minutes, the battery will not accept a
charge and should be replaced. The actual
state of the battery charge can be assessed
by using the TEST button on the battery.

Amber . . . . . . . . . . . . . . . . . . . . . . . The battery is fully charged. The rate of

charge has been reduced to a maintenance
level. The battery can be left this way
indefinitely.

Red . . . . . . . . . . . . . . . . . . . . . . . . . The receptacle (or the battery, if one is

installed) has a problem. A flashing red light
indicates that communication with the battery
is unsuccessful.

Off . . . . . . . . . . . . . . . . . . . . . . . . . . The receptacle is ready to charge a battery. If

the indicator remains extinguished when a
battery is inserted, the battery is severely
depleted. In this case, leave the battery in the
receptacle and unplug the power cord.
Reconnect the power cord and the battery
will start to charge.