HAPSITE Smart Plus Operating Manual

Diff(mmu) . . . . . . . . . . . . . . . . . . . . Provides an adjustment to DAC value for

mass peak alignment, if one is needed since
the last mass calibration check. 100 mmu =
0.1 AMU.

Peak% . . . . . . . . . . . . . . . . . . . . . . . Actual peak percentage of reference mass.

Isotope% . . . . . . . . . . . . . . . . . . . . . Percentage of the Carbon 13 isotope peak as

measured against the mass fragment.

Status . . . . . . . . . . . . . . . . . . . . . . . Status of mass peak.

OK . . . . . . . . . . . . . . . . . . . . . . . Within minimum and maximum values.

OK LOW. . . . . . . . . . . . . . . . . . . Outside of minimum value but within

acceptable tolerance.

OK HIGH . . . . . . . . . . . . . . . . . . Outside of maximum value but within

acceptable tolerance.

LOW . . . . . . . . . . . . . . . . . . . . . . Below minimum value; needs adjustment.

HIGH. . . . . . . . . . . . . . . . . . . . . . Above maximum value; needs adjustment.

FAILED. . . . . . . . . . . . . . . . . . . . Cannot located mas peak within window.

AutoTune or Manual Tune is required.

7.5.8 Right Mouse Button (RMB) Menus

In addition to the functions described above, the following menu items are available
by pressing the Right Mouse Button (RMB).

7.5.8.1 RMB in Scan Window

Placing the Mouse Cursor in the Peak Scan or Full Scan Window and pressing the
RMB displays the menu shown in

Figure 7-25

.

Figure 7-25 Scan Window Options

Undo . . . . . . . . . . . . . . . . . . . . . . . . Undo last function.