INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 525
Advertising
16 - 27
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
3
Wearing safety glasses and gloves, clean the o-ring on the VI valve with
methanol. See
Figure 16-22
.
Figure 16-22 Cleaning VI Valve O-ring
4
Clean the bottom of the manifold, where the VI valve was removed, with
methanol. See
Figure 16-23
.
Figure 16-23 Cleaning Bottom of Manifold