2 tune drop down menu – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 253
7 - 11
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
Show Bounds . . . . . . . . . . . Displays the peak centroid and the target
peak width at 10% peak height.
Save Tune . . . . . . . . . . . . . . Saves the Tune File.
Load Tune . . . . . . . . . . . . . . Loads a new Tune File and restarts tuning.
7.5.2 Tune Drop Down Menu
Figure 7-15 Tune Drop Down Menu
Mass Calibration. . . . . . . . . . . . . . . Checks and corrects the ten calibration
masses for correct location within the mass
range.
Noise Check . . . . . . . . . . . . . . . . . . Scans a "quiet" mass range (no peaks) to
determine the electronic baseline and
threshold noise level.
Perform Tune Checkup . . . . . . . . . Runs a mass calibration and noise check.
Save Tune Parameters... . . . . . . . . Saves the Tune File.
Load Tune Parameters... . . . . . . . . Loads a new Tune File and restarts tuning.
Load Factory Defaults . . . . . . . . . . Loads the default tune settings from a factory
tune file. This is intended to provide a starting
point for tuning.
Common Scale . . . . . . . . . . . . . . . . Sets all of the Mass Peak windows to the
same common scale (Y-axis), based on
Mass 117.