Show Bounds . . . . . . . . . . . Displays the peak centroid and the target

peak width at 10% peak height.

Save Tune . . . . . . . . . . . . . . Saves the Tune File.

Load Tune . . . . . . . . . . . . . . Loads a new Tune File and restarts tuning.

7.5.2 Tune Drop Down Menu

Figure 7-15 Tune Drop Down Menu

Mass Calibration. . . . . . . . . . . . . . . Checks and corrects the ten calibration

masses for correct location within the mass
range.

Noise Check . . . . . . . . . . . . . . . . . . Scans a "quiet" mass range (no peaks) to

determine the electronic baseline and
threshold noise level.

Perform Tune Checkup . . . . . . . . . Runs a mass calibration and noise check.

Save Tune Parameters... . . . . . . . . Saves the Tune File.

Load Tune Parameters... . . . . . . . . Loads a new Tune File and restarts tuning.

Load Factory Defaults . . . . . . . . . . Loads the default tune settings from a factory

tune file. This is intended to provide a starting
point for tuning.

Common Scale . . . . . . . . . . . . . . . . Sets all of the Mass Peak windows to the

same common scale (Y-axis), based on
Mass 117.