4 part 3: install the new neg pump – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 527

Advertising
background image

16 - 29

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

7

Place the HAPSITE on the Service Module and lock down with the black side
latches. Refer to

section 2.8.2, Placing the HAPSITE on the Service Module,

on page 2-24

for additional instructions, if needed.

8

Continue on to

section 16.7.4, Part 3: Install the New NEG Pump, on page

16-29

.

16.7.4 Part 3: Install the New NEG Pump

WARNING

Gloves must be worn while performing all of the
following steps. Wear safety glasses and appropriate
PPE according to the MSDS when handling chemicals.

1

Wearing safety glasses and gloves, clean the manifold flange with methanol on
a lint free wipe. See

Figure 16-26

.

Figure 16-26 Cleaning Manifold Flange