Figure 9-26 – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 331
Advertising
9 - 23
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
Figure 9-26 RIC Plot for Suspected Benzene One Mass Selected
10
The compound may be present in the unknown sample if all three masses
(peaks) align in the RIC plot. Use F7 NIST program to confirm identification of
the suspected compound. See
Figure 9-27
.
NOTE: There may or may not be a peak visible in the TIC plot.
Figure 9-27 RIC Plot to locate Benzene — Three Masses Selected
11
The compound was not detected in the unknown sample if all three masses
(peaks) are not present, or do not align together in the RIC plot.