3.10.2 Quick Reference SOP —

GC/MS Mode with SituProbe in Portable Mode

1

Verify that the desired sampling configuration (i.e. concentrator) is installed.

2

Attach the Transfer Line.

3

If the system is Shutdown or in Extended Standby, either power on the
HAPSITE or take the system out of Standby. If a Concentrator is installed, the
HAPSITE will begin preparing a Concentrator method. See

Section 3.4.3,

Procedure for Running Concentrator Methods, on page 3-25

.

3a

If the SYSTEM IS NOT READY message is displayed, touch PREPARE
ANALYZE
. Alternately, use the arrow keys to highlight PREPARE ANALYZE
and push OK SEL.

4

Optional: Fill vessel to the top of the outlet connection.

5

Place probe in vessel or water source.

6

Touch RUN ANALYZE or push ANALYZE RUN when the SYSTEM READY
screen is displayed.

7

Remember: when the screen prompts, Collecting Sample Now For and
Sampling Is Done, no action is required from the user.

8

When the run is complete, a METHOD FINISHED prompt will appear.

9

Refer to

Section 3.5.1, View Results/View Reports, on page 3-32

for

information on data review.

CAUTION

The Concentrator feature has increased sensitivity. Take
care to avoid saturating the HAPSITE.