HAPSITE Smart Plus Operating Manual

Writing a quantitative method still begins with Method Editor, but certain buttons will
not be available until the library is created. All the settings described in

section

11.9.1, Setting Up a Qualitative Search, on page 11-36

still apply to the creation of

a quantitative method. In addition two more options become active, The Library
Search Parameters button and the Library Actions window, see

Figure 11-36

.

Figure 11-36 Library Search Parameters Button

Clicking the Library Search Parameters button opens the window shown in

Figure

11-37

. The center section refers to the AMDIS parameters and is exactly the same

as described in

Figure 11-28 on page 11-37

.

Figure 11-37 Library Search Parameters Window

The Peak Search section is comprised of all the elements that define a peak in the
chromatogram.

Search Window . . . . . . . . . . . . . . . This value indicates the time window allowed

for a peak to appear. 30 seconds indicates
that the acceptable window for a peak is 15
seconds on either side of the expected
retention time.

Window Expand Factor . . . . . . . . . This value has a default of 0.05. This value is

multiplied by the run time and then added to
the search window time. The search window
is then expanded by the time calculated.