2 gc temperature profile – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 382

Advertising
background image

11 - 22

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

SPConcFill . . . . . . . . . . . . . . . . . . . This sets the GC valves so that the sample

pump pulls the sample through the
concentrator.

SPLoopFill . . . . . . . . . . . . . . . . . . . This sets the GC valves so that the sample

pump pulls the sample through the Sample
Loop. The recommended duration is at least
60 seconds to allow for clearing the sample
line of the previous sample as well as good
mixing with the Internal Standard (if used).

SPN2DryPurge . . . . . . . . . . . . . . . . This Inlet State purges the transfer line and

concentrator system with dry N

2

before

sample injection to remove moisture.

11.5.2 GC Temperature Profile

WARNING

Only trained users should modify methods. Changing GC
temperature profiles may result in incorrect analysis.

GC Temperature Profiles (see

Figure 11-18

) specify the temperature ramp and

hold settings for the HAPSITE Method. Any profile can be edited by changing the
value in the Temp, Ramp Rate or Hold Time column. When a new profile is added,
the first value entered will force the other values to be calculated and entered
automatically. However, this automatic entry does not prohibit the values from
being edited. The Start time of each profile is automatically derived from the
preceding profile. The Ramp Time is automatically calculated based on the Ramp
Rate
and Temp entries. It cannot be edited directly.

Figure 11-18 Method Editor Inlet Page: GC Temperature Profiles

The Insert key does not work for this grid. Changes to the profiles must be made
by editing the Temp, Ramp Rate or Hold Time and adding heater profiles.