3.3.1 Quick Reference SOP — Survey Method

1

If you are running an Analyze (GC/MS) method after Survey, verify that the
appropriate configuration (i.e., concentrator) is installed and the proper Analyze
method is displayed on the screen.

2

If powering on HAPSITE or ending Extended Standby, the HAPSITE will
automatically begin preparing Survey.

2a

If needed, touch PREPARE on the touch screen.

2b

Alternately, using the arrow keys, highlight PREPARE. Push OK SEL.

3

When prompted by the SYSTEM READY message, touch RUN SURVEY or
push SURVEY RUN.

4

Monitor background for one minute.

5

Hold the probe over the sample.

6

Remove the probe from the sample when the TIC begins to increase and a
peak begins to form.

6a

If the TIC does not increase after a full minute of sampling, remove the probe
from the sample.

7

To confirm data with an Analyze (GC/MS) run, touch RUN ANALYZE or press
ANALYZE RUN while sample is being detected.

8

If an Analyze method is not desired, touch STOP or push SURVEY RUN.

9

A METHOD FINISHED message will be displayed when the method has
ended.