2.10.4 Removing the Concentrator
1
Open the front panel of the HAPSITE and remove the black cover labeled
CONCENTRATOR.
WARNING
The elbow fittings, nuts and Concentrator may be hot.
Allow these components to cool before continuing.
WARNING
Excessive force and/or tightening can cause the fragile
glass to break!
2
With fingers, loosen the nuts on the top and bottom of the Concentrator until the
it becomes free.
3
Lift the top elbow. Gently lift and angle the Concentrator out of the fixture.
4
Remove the Concentrator from the bottom elbow, being careful not to lose the
ferrules inside the nuts.
5
Store the Concentrator wrapped in tissue in its storage vial for future use.