INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 512

Advertising
background image

16 - 14

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

9

Install the red probe cap (P/N 059-0382). See

Figure 16-10

.

10

Install the brass probe nozzle cap. See Figure 15-10.

Figure 16-10 Red Probe Line Cap and Brass Probe Cap

NOTE: The outside of the probe line can be decontaminated while it is still

attached to the Analytical Module. To decontaminate the inside of the
probe line, refer to

Contamination of the Probe and Probe Line, section

16.5.3 on page 16-9

CAUTION

All caps must be in place and the front door must be
firmly shut before decontaminating the Analytical
Module.

WARNING

Make sure that you are in a well ventilated area.

WARNING

Wear appropriate PPE (Personal Protective Equipment).
Refer to the MSDS for the appropriate PPE.

11

Wipe the Analytical Module down using a soft cloth and a 10% hydrochlorite
(bleach) solution (i.e., 1 part bleach to 9 parts water).

12

Use a soft cloth and warm water and wipe the Analytical Module down again.

Advertising