INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 488
15 - 14
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
3
Double-clicking on the Service Module icon will open the Service Module tab
in the HAPSITE Properties window, as shown in
Figure 15-15
.
Figure 15-15 Service Module Tab Selected in HAPSITE Properties Window
4
In the Service Module tab shown above, select the Attach button. A
confirmation window will open. Click Yes. See
Figure 15-16 Attach Confirmation Request
5
The Roughing Pump will start first, then the Turbo Pump will begin to speed up,
as shown on the Turbo Speed (Hz) line in
Figure 15-15
above (initially listed as
0).
NOTE: The Attach procedure typically takes about 5 minutes to complete.
After selecting the Attach button, the HAPSITE Properties window can be
closed at any time.
While attaching, the following prompt will appear.
Figure 15-17 Attach In Process