4 tune files – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 350

Advertising
background image

10 - 4

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Figure 10-4 Method Editor - Data, Showing File Name Options

The system allows flexibility to add Date or Date and Time as a prefix or suffix. A
different naming convention if Use Default Filename is not selected.

10.1.4 Tune Files

Refer to

Chapter 7, Tune

for additional information on Tune.

NOTE: Manual Tuning requires Advanced access. Normal users do not have

access to Manual Tune. Refer to

section 8.9, Access Levels, on page 8-23

for additional information on Access Levels.

If Manual Tune is performed, the file must be saved to the Tune folder on the
HAPSITE.

Figure 10-5

shows a typical Manual Tune screen, which is opened by

selecting the Tune icon in the System Setup window, or by right-clicking on a
sensor icon and selecting Tune. To save changes made during manual tuning,
press the Save Tune button on the Sensor Toolbar and tune will be saved
immediately. Other ways to save a tune file include using File >> Save, File >>
Save As
, or Tune >> Save Tune Parameters.