HAPSITE Smart Plus Operating Manual

Figure 2-35 System Setup Screen with Service Module

NOTE: If the Service Module icon does not appear, the communications may need

to be configured. See

section 8.6.3, Establishing Communication with the

Service Module, on page 8-20

, for information on how to set up a COM port

for communicating with the Service Module.

Click on the Vent icon. The dialog shown in

Figure 2-36

will appear to confirm the

venting of the Service Module.

Figure 2-36 Vent Service Module Confirm Window

Select Yes to vent the Service Module. A window will appear counting down a
fifteen-second delay while the vent procedure completes. Once completed, remove
the aluminum plug.

Next, clean any debris or dust from the Viton® o-ring using a lint-free wipe with
methanol. Wipe the top of the o-ring, following the contour to clean the entire top
exposed section. Avoid pushing dust or debris into the middle opening. Cleaning
will ensure a tight seal to the HAPSITE, allowing the system to maintain vacuum.
See

Figure 2-37

, which shows the o-ring being cleaned.

WARNING

Wear gloves and safety glasses when handling methanol.