INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 217

5 - 11
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
4
The HAPSITE will check pressures and automatically heat up all necessary
zones to the setpoint temperatures specific to the selected method. Progress
will be indicated by a bar graph.
5
If this method is not the desired method, see
6
Click the RUN METHOD icon. Click on the GC/MS folder. Click on the folder of
the desired configuration and then click on the method folder. See Steps 2-4 of
.
7
Click RUN button on the pop-up window or from the control panel on the Laptop
screen. See
.
Figure 5-15 Run Button
8
When the HAPSITE screen prompts Collect Sample Now, place the probe
over sample for the specified amount of time.
CAUTION
Do not touch the sample with the probe. Do not place the
sample Probe in liquids while sampling.
9
Move the probe away from the sample when the screen prompts Sampling
Done.
10
As the method runs, the chromatogram will begin to appear on the screen. See
Figure 5-16
for an example of a chromatogram in progress.