INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 429

Advertising
background image

12 - 11

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

2

After the standard(s) have been prepared, a method must be developed to
collect the data. The method can be a default method or a modified method
using the Method Editor. All method development and chromatographic
changes which can affect the retention time must be made prior to collecting
the standard runs that are to be used for building and calibrating the library.
Refer to

Chapter 11, Method Editor

for additional information on method

development.

HINT: When running each sample, enter the concentration and a description

on the Data File Information page. See

Figure 12-9

. To access this page

refer to

section 9.3, Data Review Toolbar, on page 9-5

.

Figure 12-9 Data File Information Page

NOTE: In this example, a series of standards at four concentration levels were

prepared by serial dilution. The method, QuantExample, was used to
collect the data. Start with one method for collection and continue with
that method. These are arbitrary concentrations for example only and
NOT actual concentrations.

QuantExample_20060301_005 = 150 PPM

QuantExample_20060301_004 = 50 PPM

QuantExample_20060301_002 = 200 PPM

QuantExample_20060301_001 =100 PPM

3

After the data has been collected, open the Calibration function which can be
accessed from the drop-down menu, by right mouse button clicking on the
HAPSITE icon, by the toolbar icon or via the Status icon and Function tab.
See

Figure 12-10

.

Advertising