10 situprobe, 1 procedure for situprobe operation – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 167

Advertising
background image

3 - 73

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

3.10 SituProbe

For complete information on SituProbe methods, see

Chapter 14, SituProbe

. For

instructions on setting up the SituProbe, refer to

Section 2.9, SituProbe, on page

2-27

.

3.10.1 Procedure for SituProbe Operation

1

Verify that the desired sample configuration is installed (i.e., Concentrator).

2

Make sure that the Transfer Line is attached.

3

When powered on or taken out of Extended Standby, the HAPSITE will
automatically start preparing a SituProbe method. If the SituProbe method that
the HAPSITE begins preparing is not the desired one, refer to

Section 3.2,

Selecting a Different Method Using the SELECT METHOD Icon, on page 3-13

.

4

The HAPSITE will display the progress of its preparation. Refer to

section 3.1

,

Step 5 for more information.

5

If using the water vessel, fill the vessel to the top of the outlet connection. See

Figure 3-119

.

Figure 3-119 SituProbe Vessel

6

Insert probe into vessel or water source.

6a

If running a SituProbe method that requires a Concentrator, the HAPSITE will
run a Concentrator Cleanout. Its progress will be shown by a bar graph. If the
Concentrator Cleanout is not successful, refer to

section on page 3-29

.

7

When the HAPSITE has finished preparing, a SYSTEM READY message will
appear with a prompt to press Survey or Analyze to begin sampling.

NOTE: A blank run is recommended before running a sample.

Top Of Outlet

Connection

Advertising