INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 227
Advertising
5 - 21
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
10
A prompt will appear Press RUN to start method.
NOTE: The sample must be heated for 20 minutes prior to sampling.
11
Click RUN button on the pop-up window or from the control panel on the
screen.
Figure 5-27 Run Button
12
The HAPSITE will prompt Collecting Sample Now. The Headspace will
automatically start sampling. No action is required by the user.
13
When the HAPSITE screen prompts Sampling Done, the Headspace has
finished sampling. Again, no action is required by the user.
Figure 5-28 HSS Sampling Done
14
The chromatogram, showing the run’s progress, will appear on the Laptop
screen. See
.