INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 526

Advertising
background image

16 - 28

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

5

Replace the VI valve into the bottom of the manifold. Hand tighten.
See

Figure 16-24

.

Figure 16-24 Replacing the VI Valve

6

Clean the o-ring on the Service Module with Methanol on a lint free wipe. See

Figure 16-25

.

Figure 16-25 Cleaning O-ring on the Service Module

Advertising