HAPSITE Smart Plus Operating Manual
5
Replace the VI valve into the bottom of the manifold. Hand tighten.
See
Figure 16-24
.
Figure 16-24 Replacing the VI Valve
6
Clean the o-ring on the Service Module with Methanol on a lint free wipe. See
Figure 16-25
.
Figure 16-25 Cleaning O-ring on the Service Module