HAPSITE Smart Plus Operating Manual
5.7.2 Quick Reference SOP — Analyze Mode with SituProbe
1
Double-click the Plus IQ software icon.
2
Verify that the desired sample configuration (i.e., Sample Loop or
Concentrator) is installed.
3
Make sure the SituProbe is set-up properly. See
for assembly information.
4
If necessary, power on the HAPSITE or take the system out of Extended
Standby.
5
Depending upon the method, the HAPSITE will begin preparing the
Concentrator. If needed, see
Section 3.1.2, Concentrator Options (CONC
Section 5.3, Selecting a New Method, on page
if the start up method is not the desired method
6
When the system has finished preparing, click the RUN METHOD icon. Select
the desired method from the appropriate folder. Refer to Steps 2-4 of
7
Click the Run button in the pop-up window.
8
When the screen prompts, Collect Sample Now and Sampling Is Done, no
action is required from the user.
9
When the run is complete, a Method Finished prompt will appear.
10
To analyze the data, see