4 operating specifications – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 461

Advertising
background image

14 - 3

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

14.4 Operating Specifications

Environmental Temperature . . . . . . . Operating: 5 to 45°C, 0-95% RH

(non-condensing)
Storage: -40 to 70°C (stored dry)

Water Sample

Temperature . . . . . . . . . . . . . . . . 5 to 30°C

Above 26°C, Dry Purge may be extended

Conductivity . . . . . . . . . . . . . . . . ionic conductivity is used to sense the proper

water level in the probe head before starting
the analysis. Therefore, the water sample
must exhibit conductivity before the analysis
can begin. Very pure water, such as
deionized or distilled water, may require the
addition of a salt before the analysis
sequence will start.

pH Range . . . . . . . . . . . . . . . . . . 2-11

Probe Depth . . . . . . . . . . . . . . . . . . . minimum, 5.5 in (140 mm)

maximum, 18 in (457 mm)

Heated Zones

Probe . . . . . . . . . . . . . . . . . . . . . 30 to 60°C

Oven . . . . . . . . . . . . . . . . . . . . . . 30 to 60°C

Transfer Line. . . . . . . . . . . . . . . . 30 to 60°C

80°C during Flush Mode
(External Power Only)

Practical Quantitation Limit. . . . . . . . ppt to ppb, based on HAPSITE configuration

and method parameters

Internal N2 Carrier Gas: . . . . . . . . . . Standard HAPSITE (770 kPa) N2

disposable can

Power Requirements

DC Power . . . . . . . . . . . . . . . . . . 24 V(dc) ±5% via HAPSITE Power Adapter

80 W maximum

Battery Power . . . . . . . . . . . . . . . Standard 19.2 V(dc) HAPSITE battery. Flush

mode not recommended on battery power.

Software . . . . . . . . . . . . . . . . . . . . . . HAPSITE Plus IQ Software

Advertising