HAPSITE Smart Plus Operating Manual

NOTE: Clean any debris or dust from the black rubber o-ring using a lint-free wipe.

Wipe the top of the o-ring, following the contour to clean the entire top
exposed section. Avoid pushing dust or debris into the middle opening
leading to the pumps.

Cleaning the o-ring will help ensure that a good seal to the aluminum storage plug
is achieved. The seal allows the system to maintain vacuum more effectively. See

Figure 15-29

shows the o-ring being cleaned.

Figure 15-29 Cleaning the Service Module Rubber O-ring using a Lint-free Wipe

After cleaning the rubber o-ring, make sure there are no cuts. Also look for any
visible cracking of the o-ring. If the o-ring is damaged, it may need to be replaced.

To place the Service Module in storage, the aluminum storage plug is placed on the
opening where the HAPSITE connects, as shown in

Figure 15-30

.

Figure 15-30 Service Module Aluminum Storage Plug in Place

HINT: Confirm the RS-232 communication cable is connected to the Service

Module and Laptop. If the RS-232 communication cable is not attached,
refer to

section 2.8.1, Setting Up the Service Module, on page 2-19

before

proceeding.