HAPSITE Smart Plus Operating Manual

Figure 5-20 AutoTune Complete

5

A Concentrator Cleanout will also be run as part of the HAPSITE’s preparation.
If the Cleanout is successful, a green SYSTEM READY message will be
displayed. If the Cleanout is unsuccessful, refer to

Section 3.1.3, Concentrator

Cleanout Failure, on page 3-11

. If this method is not the desired method, refer

to

Section 5.3, Selecting a New Method, on page 5-7

.

6

Click the RUN METHOD icon. Click on the GC/MS folder. Click on the folder of
the desired configuration and then click on the method folder. See Steps 2-4 of

section 5.3 on page 5-7

.

7

Once all temperature zones have reached their setpoints, a prompt will appear
Press RUN to start method.

8

Click RUN button on the pop-up window or from the control panel on the
screen.

Figure 5-21 Run Button

9

When the HAPSITE screen prompts Collect Sample Now, sample for the
specified amount of time. See

Figure 5-22

.

Figure 5-22 Collecting Sample for Concentrator