HAPSITE Smart Plus Operating Manual
14
At the end of the Headspace run, a prompt will appear. When prompted Insert
needle into a clean vial and press Continue to Purge, place the needle into
the empty vial. When finished, touch the Continue button or using the arrow
keys, highlight Continue. Push OK SEL.
Figure 3-117 Purge
NOTE: If purge is skipped, a full Headspace run must be completed to flush
the system. Once purge has been skipped, it is not possible to
re-access it.
15
The METHOD FINISHED message will appear when the purge method has
ended. See
Figure 3-118
.
Figure 3-118 Sample Method Finished for HSS
NOTE: Another Analyze (GC/MS) method can be started immediately after one
has completed. Depending on the temperature profile, the column may
need to cool before another run will begin.
NOTE: Refer to
Section 3.5.1, View Results/View Reports, on page 3-32
information on reviewing the data.