3.7.1.4.1 HEATERS Button

Touching the HEATERS icon provides the current temperatures of the column, the
membrane, the GCHL, the valve oven, the NEG, the probe, and the Concentrator
Elbow as the HAPSITE is heating. The number after the actual temperature is the
setpoint temperature. A green circle signifies that the temperature is in range, blue
signifies that the heater is reaching its setpoint, and black signifies that the heater
is off. See

Figure 3-88

.

Figure 3-88 Temperatures of Heaters

3.7.1.4.2 NEG Button

The NEG button provides information about the NEG. This includes the NEG’s
current and setpoint temperatures, the On Time which is the number of hours that
the NEG has been used, and the date the NEG was activated. There is also
information about the number of times the NEG has been activated and the date(s)
of reactivation.

From this screen, there is a button to activate the NEG and also a button for NEG
bakeout. See

Figure 3-89

.

Figure 3-89 NEG Button