2 using the hapsite with hss, 1 powering on the hss – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 444
13 - 6
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
13.2 Using the HAPSITE with HSS
The following instructions can be used as a guideline for sample and standard
preparation unless more specific operating instructions are available.
13.2.1 Powering on the HSS
CAUTION
Do not open the front door in a contaminated or wet area.
Open the front door and press the POWER button to turn on the HSS. See
Figure
13-3
.
HINT: The POWER button is a toggle switch. Once pressed and released the
switch returns back to its previous position.
Figure 13-3 Powering on the Headspace Instrument
,
Once the HSS is connected to the HAPSITE, and the HSS power is on, run one of
the Headspace methods using the Run Method icon. This will heat up the
Headspace instrument. See
section 5.6, Analyze (GC/MS) Mode with Headspace
section 3.9, Analyze (GC/MS) Mode with