INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 210

Advertising
background image

5 - 4

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

10

Monitor the Probe distance indicator bar on the HAPSITE screen for guidance.
See

Figure 5-5

. The bar rises as the TIC increases. To avoid saturation, remove

the Probe from the sample when the bar is high and turns yellow.

Figure 5-5 Survey Probe Distance Indicator Bar

11

Move the probe away from the sample of interest. Monitor the TIC count until
the level returns to the initial background level that was noted in step 6.

12

Click on the STOP button (in the center of the control panel on the right side of
the screen) to stop the sampling process and save the data. See

Figure 5-6

and

Figure 5-7

.

Figure 5-6 Stop Button

NOTE: Remember that this is a tentative identification. For a more positive

identification, run an Analyze (GC/MS) method for confirmation of the
results.

Advertising