1 how to change or remove a gas canister, Warning, Caution – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 47
2 - 9
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
2.4.3.1 How to Change or Remove a Gas Canister
Push the lever located on the right of the canister. The canister will release. (A
slight twist of the canister may be required.) Remove the canister. Refer to
Figure
2-16 on page 2-9
.
NOTE: The Nitrogen canister will need to be replaced after roughly 8 hours of use.
The Internal Standard canister will need to be replaced after 3 to 4 days of
continuous use. These numbers are guidelines and will vary.
WARNING
Do not refill the canisters. Bodily injury may result.
Canisters are designed to be disposable and may fail if
filling is attempted.
CAUTION
Closing the front panel when the canisters are not
properly installed may damage the HAPSITE and/or
canisters.
Figure 2-16 Canister Removal
Carrier Gas
Canister
Internal
Standard
Canister
Press
Lever