2 reactivating the neg pump, 1 reactivation using the laptop – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 517
Advertising
16 - 19
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
16.6.2 Reactivating the NEG Pump
Reactivating the NEG Pump requires having the Service Module attached to the
HAPSITE for at least 22 hours. The procedure is basically the same as activating
a new NEG Pump. Refer to
section 2.8, Service Module, on page 2-19
information on setting up the Service Module.
16.6.2.1 Reactivation Using the Laptop
1
Click on the Service Module tab of the Properties menu.
2
Click on the Service button. See
Figure 16-16 Service Button on the Service Module Tab
3
Use the Activate NEG settings below as a guideline. However, change the
HAPSITE Pumpdown Time which should be set to 180. See
Advertising