6 part 5: activation of the neg pump – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 538

Advertising
background image

16 - 40

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

16.7.6 Part 5: Activation of the NEG Pump

1

Click on the Service Module tab of the Properties menu.

2

Click on the Service button. See

Figure 16-46

.

Figure 16-46 Service Button on the Service Module Tab

3

Use the Activate NEG settings below as a guideline.

HAPSITE Pump Down Time (min) = 60

NEG Activation Power (Watts) = 30

Activation Time (min) = 480

Activation Cool Down Temp (C) = 400

Activation Cool Down Time (min) = 720

See

Figure 16-47

.

Figure 16-47 Service Window