6 part 5: activation of the neg pump – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 538
Advertising
16 - 40
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
16.7.6 Part 5: Activation of the NEG Pump
1
Click on the Service Module tab of the Properties menu.
2
Click on the Service button. See
Figure 16-46 Service Button on the Service Module Tab
3
Use the Activate NEG settings below as a guideline.
HAPSITE Pump Down Time (min) = 60
NEG Activation Power (Watts) = 30
Activation Time (min) = 480
Activation Cool Down Temp (C) = 400
Activation Cool Down Time (min) = 720
See
Figure 16-47 Service Window