HAPSITE Smart Plus Operating Manual

14.5 Consumables

14.5.1 Compressed Nitrogen

A source of pressurized, ultra high purity Nitrogen is required for sample purging
and analysis. Nitrogen delivered by on-board canister or an external Nitrogen
source, regulated to 60-100 psi (approx. 420-700 kPa), may be connected through
the 3/16 in Swagelok quick-connect fitting located at the back of the instrument.

External ultra high purity Nitrogen cylinders must meet the following requirements:

less than 50 ppb VOCs

less than 40 ppm Argon

The default methods for the SituProbe consume approximately 370 mL of Nitrogen
per run. Therefore, approximately 8 samples can be collected from a full Nitrogen
Gas canister. For mobile, in-the-field sampling, a 110 L bulk tank of Nitrogen is
available. Larger bulk cylinders of ultra high purity Nitrogen, e.g., 300 ft

3

, can be

used in fixed locations.

WARNING

Nitrogen gas canisters are under pressure.