INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 404

Advertising
background image

11 - 44

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

The Reset button allows for default values to be reset in the Peak Search window.
See

Figure 11-38

.

Figure 11-38 Resetting Default Search Parameters

The Elimination Rules section gives parameters for peaks to be reported. There
are three options. See

Figure 11-39

.

Figure 11-39 Elimination Rules Window

Keep no more than. . . . . . . . . . . . . This check box determines how many peaks

are to be shown. This refers to the Mass
Spectrum. If 10 is selected, only the top ten
peaks will be displayed.

Keep peaks above . . . . . . . . . . . . . This check box defines the percent intensity

of a mass peak to be displayed on the Mass
Spectrum. Fifteen is generally used as the
low end for intensity.

Maximum NIST Hits . . . . . . . . . . . . This value is the number of matches reported

by NIST.

NOTE: This section also has a Reset button, which will set any changed values

back to the default setting.

The Library Action box has four main sections. Each can be enabled using a
check box. See

Figure 11-40

.

Figure 11-40 Library Actions

AMDIS . . . . . . . . . . . . . . . . . . . . . . . When enabled, a concentration can be

entered. This will determine the critical level
of concentration at which AMDIS will identify
the chemical on the front panel display of the
HAPSITE.