HAPSITE Smart Plus Operating Manual

11

Spray the MicroDuster around the orifices listed below that could be the source
of a possible leak, while checking the scan on the laptop in the tune program.
If there is a leak, the response for the 69 and 83 masses will increase. See

Figure 16-44

.

The critical areas to be checked are:

The seam between the NEG Pump and the manifold.

The seam between the Service Module and Analytical Module on the
manifold side.

Figure 16-44 Spraying MicroDuster

12

After the manifold has been leak tested and no leaks found, plug the black
cable on the front of the NEG Pump into the black socket located above the
NEG Pump.

13

In the Tune window, turn the IS Supply and Tune Feed on by clicking on the
buttons which should change from the color black to the color green. See

Figure 16-39

.

14

At this point, a tune check should be performed on the system. Click on the
Short Tune icon. See

Figure 16-45

. Refer to

section 7.2, AutoTune, on page

7-2

.

Figure 16-45 Short Tune Icon

15

Continue on to

section 16.7.6, Part 5: Activation of the NEG Pump, on page

16-40

.