HAPSITE Smart Plus Operating Manual

12

To view the chromatogram while the method is running, touch GRAPH. See

Figure 3-39

. Alternately, use the arrow keys to highlight GRAPH and push

OK SEL.
NOTE: This screen will also display the TIC Max, the current TIC and the time

the method has been running.

NOTE: Touching the blue compound identification above the chromatogram

will display its NIOSH database information. See Step

8 on page 3-35

.

Figure 3-39 Sample GRAPH function

13

When the TIC begins to increase, move the probe away from the sample of
interest. Continue the run until the TIC level returns to the initial background
level that was noted in Step

8 on page 3-20

. See

Figure 3-40

.

NOTE: Monitor the Probe distance indicator bar on the HAPSITE screen for

guidance. The bar rises as the TIC increases and green signifies that
the proper sampling distance is being maintained. To avoid saturation,
remove the probe from the sample when the bar is high and turns
yellow. If saturation occurs, the side bar will turn red and the TIC will be
above 60 million.

Remove the probe from the sample and continue to

run Survey until the saturation is cleared from the sample pathway.

Figure 3-40 Returning to Baseline