Figure 2-63, Warning – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 78
2 - 40
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
Figure 2-63 Installed Concentrator Cover
NOTE: If the cover does not easily fit over the Concentrator, do not force it on.
Check to ensure the Concentrator is correctly installed with the
Concentrator fully seated into both elbows and the nuts properly
tightened for a secure fitting.
NOTE: The Concentrator Cover contains two metal contacts that are
supported by Fiberglass. Inspect the contacts, prior to assembly, to be
sure the contacts are not bent or crimped. See
.
WARNING
Use the proper Concentrator cover, which contains an
Electronic Circuit board, for the Smart Plus. Do not use
the old Concentrator cover from the HAPSITE Smart
Plus. If the old cover is used, the cover will overheat and
may become damaged.
Figure 2-64 Back of the New Concentrator Cover