HAPSITE Smart Plus Operating Manual
11.8.1 SIM for Analyze (GC/MS)
Figure 11-25 Method Editor SIM Page for Analyze (GC/MS) Analysis
SIM for the Analyze (GC/MS) mode allows for creation of multiple sets of masses.
Each set has a Begin Time and an End Time which must be entered when
programming the Set. After the Begin and End Times are entered, as well as an
optional Name, the Mass list for the Set can be entered. Default settings for each
Mass are automatically entered based on the previous Mass. As the Mass list is
entered, the Round Trip Time is automatically calculated and entered in the Scan
Sets grid.
The Scan Sets fields are as follows, in the order recommended for editing:
Begin Time . . . . . . . . . . . . . . . . . . . In minutes: seconds, the time at which
collection of the listed masses should start.
End Time . . . . . . . . . . . . . . . . . . . . . In minutes: seconds, the time at which
collection of the listed masses should stop.
Name . . . . . . . . . . . . . . . . . . . . . . . . Each Scan set can be assigned a name for
identification purposes. This entry is optional.