INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 274
Advertising
7 - 32
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
3
After making Ion Energy adjustments:
3a
If the red boxes no longer appear, indicating the Actual Resolution is within
range, proceed to
Step 4
. See
.
3b
If the HAPSITE is failing to Tune, then repeat the Manual Tune procedure.
Proceed to
section 7.6.1, Adjusting Base Peak Gain, on page 7-26
. If this
procedure is repeated more than 2 times, contact INFICON.
Figure 7-36 Good Manual Tune
4
Save the Tune file. Click Save Tune button on the upper Toolbar. See
Figure
7-37
and
.
93
90
170
117 140 170
167
140
220
213 175 230
246 180 230
263 185 250
282
190 255
Table 7-1 Manual Adjustments Guidelines (continued)
MASS
IE low
IE high