5.6 Analyze (GC/MS) Mode with Headspace Sampling System

The Headspace Sampling System is used to test liquid and soil samples. See

Chapter 13, Headspace Sampling System

for more information on the Headspace

System. Refer to

Section 2.7, Headspace Sampling System, on page 2-13

for

assembly instructions. See

Chapter 6, Methods

for additional information on

GC/MS methods.

When used with the Concentrator, the Headspace can analyze samples with
concentration levels in the low part per billion to high part per trillion range. It is used
when the Headspace Sampling System with the Sample Loop method fails to
detect a chemical. See

Chapter 6, Methods

for additional information on Analyze

(GC/MS) methods.

CAUTION

The Concentrator feature has increased sensitivity. Take
care to avoid saturating the HAPSITE.

1

Verify that the desired sample configuration is installed (i.e., Sample Loop or
Concentrator.)

2

Double-click the Plus IQ icon on the Laptop screen.

3

Make sure that the Headspace Transfer Line is installed.

4

Place a 40 mL vial with either blank or sample into a Headspace well.

NOTE: Do not fill 40 mL vial more than 20 mL full. See

Section 13.2.3, Loading

the Wells, on page 13-11

for detailed instructions about loading the

sample vials into the wells.

5

Place a clean, empty 40 mL vial into a Headspace well.

WARNING

The Headspace needle is very sharp!