3 vx / r-33 conversion tube, 1 vx / r-33 conversion tube installation – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 82

Advertising
background image

2 - 44

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

2.11.3 VX / R-33 Conversion Tube

This procedure describes the steps required to prepare the HAPSITE to sample for
VX or R-33 using the Conversion Tube. To detect VX or R-33, you must insert the
Conversion tube following the guidelines below.

The process of detecting VX or R-33 on the HAPSITE requires the conversion of
VX or R-33 (high boiling point chemicals) to the G analog. The VX or R-33 molecule
is broken at the sulfur bond when it comes in contact with a silver fluoride pad. The
result is the formation of a volatile chemical ethyl methyl phospho no fluoridate in
the case of VX, or isobutyl methyl phospho no fluoridate in the case of R-33. These
compounds are detected by the HAPSITE as VX-G or R-33-G.

NOTE: Theoretically, other G agents can be detected with the VX conversion pad

in place. However, if other G agents are suspected, the sample should also
be run without the conversion tube in place.

WARNING

Sulfur mustard cannot be detected with the conversion
tube in place.

2.11.3.1 VX / R-33 Conversion Tube Installation

1

Refer to

section 2.11.1, Probe Nut Assembly, on page 2-42

to ensure proper

ferrule orientation in the probe nut.

WARNING

Correct ferrule placement is critical to ensure a leak-free
fit around the conversion tube.

2

Loosen the nut on the end of the probe approximately 1/4 to 1/2 of a turn. See

Figure 2-67

.

Figure 2-67 Loosening the Nut

Advertising