2 performance specifications – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 30

Advertising
background image

1 - 2

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Headspace Sampling System . . . Also known as the HSS, the HSS is an

accessory to the HAPSITE that allows testing
for volatile compounds in solids and liquids,
including soil and water.

SituProbe™. . . . . . . . . . . . . . . . . . . The SituProbe accessory is a water purging

device that provides continuous testing of
water samples in the field.

1.2 Performance Specifications

The performance specifications for the HAPSITE are shown below:

Mass range . . . . . . . . . . . . . . . . . . . . . . . 1-300 AMU

Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . as much as 1000 AMU/sec @

10 points per AMU

Ionization Mode . . . . . . . . . . . . . . . . . . . . 70 eV electron impact

Vacuum System . . . . . . . . . . . . . . . . . . . . 15 l/sec NEG pump;

0.2 l/sec sputter-ion pump

Operating temperature range . . . . . . . . . . 0

o

C to 45

o

C (32

o

F to 113

o

F)

Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . .95% RH, non-condensing

Dimensions (LxWxH) . . . . . . . . . . . . . . . . 46 x 43 x 18 (cm); 18 x 17x 7 (in)

Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Kg (35 lb) without the battery

Internal Power Consumption . . . . . . . . . . 30 watts average, 24 V(dc)

Carrier Gas . . . . . . . . . . . . . . . . . . . . . . . Nitrogen

Column Temperature Range . . . . . . . . . . 60

o

C to 180

o

C

Maximum Sample Moisture Content. . . . . 8% by weight

pH Range of Sample . . . . . . . . . . . . . . . . 2 - 11

Boiling Point of Sample . . . . . . . . . . . . . . <250

o

C

Chemical Composition of Sample . . . . . . 1 - 12 Carbon atoms

GC Column . . . . . . . . . . . . . . . . . . . . . . . 100% methyl silicone phase,

30 m x 0.32 mm ID x 1.0

 film

SIM Channels . . . . . . . . . . . . . . . . . . . . . 10

External Communications . . . . . . . . . . . . Ethernet Port, Wireless

Carrier Gas Use-Rate . . . . . . . . . . . . . . . 1 canister per 8 hours of operation

(This depends on the details of the
method being used.)