9.5.3.2 Additional Features of the Range Tool
If the cursor is placed over R1, and the RMB is clicked, the following menu items
are available. See
Figure 9-15 Range Tool RMB Menu
Remove . . . . . . . . . . . . . . . . . . . . . . Removes the Range cursor
Select Mass for Integration . . . . . . Selects the TIC or RIC for Integration
Show Integration . . . . . . . . . . . . . . Displays the integration on the x-axis
Show Retention Time . . . . . . . . . . . Displays the retention time on the x-axis
Show Signal/Noise . . . . . . . . . . . . . Shows Signal to Noise Ratio. A Background
Must be Selected using B1 first.
Subtract Current Spectrum . . . . . . Subtracts the current spectrum (Green
Triangle) from the range.
9.6 Using the Zoom Function
WARNING
Be careful not to leave out data/peaks that may be needed
for proper identification within the chromatogram.
The Zoom function allows targeting a particular section of the
chromatogram for a detailed view.
Co-elution of peaks can be detected by zooming in on odd shaped peaks.