HAPSITE Smart Plus Operating Manual
Compound names and CAS#’s. See
Figure 12-13
. The Chromatogram display
window will automatically open below the calibration control panel to display the
plot and the detected peaks.
The peaks marked with a T have been identified by AMDIS, and those with only
the black triangle have not been identified. All of the features described in
can be used to interact with the plot.
Figure 12-13 Calibrate Display with Detected Peaks
10
The template can now be edited to remove duplicate and unidentified entries
from the analyte list. Place the mouse cursor over the compound name of an
entry to be deleted or edited. Refer to
Figure 12-13
.
HINT: Only keep one entry for each compound. In this example, the Internal
Standards and the compound of interest are repeated multiple times.
Delete the duplicate entries.
Figure 12-14 Right Mouse Button in Calibration Control Panel
Press and hold the Right Mouse Button. The options available are: